April, 2004
Dear Fluke customer,
In cooperation with U.S. Consumer Product Safety Commission, Fluke Corporation is voluntarily recalling its TL221, TL222 and TL224 modular test leads. We are asking you to send back your test leads for replacement. This recall does not affect the test leads permanently attached to test probes that come standard with most Fluke Digital Multimeters. The recall does not include currently manufactured TL221, TL222 and TL224 test leads that are marked with a SureGrip™ symbol, either by a tag around the silicon wire, or the symbol molded onto the connector end. This recall does not affect the test leads permanently attached to test probes that come standard with most Fluke Digital Multimeters.
The recall does not include currently manufactured TL221, TL222 and TL224 test leads that are marked with a SureGrip™ symbol, either by a tag around the silicon wire, or the symbol molded onto the connector end. 
The recall only impacts the following SureGrip™ modular test lead sets: • TL221 • TL222 • TL224
The affected leads are also included with the following test lead kits: • TL220 • TL223 • TLK220 • TLK225 • L215
The affected leads are also included with the following Fluke equipment with serial numbers between 83260001 and 85210001:
| • Fluke 179/1AC |
• Kit-87-3 |
• Fluke 88 |
• Fluke 743B |
| • Fluke 179/EDA |
• Flk-87E-3 |
• Fluke 741B |
• Fluke 744 |
|
Description of problem
TL221, TL222 and TL224 SureGrip™ test lead sets, manufactured between December 2002 and October 2003, may have a faulty connection that could cause the tester to not indicate that voltage or current is present. THIS COULD PLACE THE USER IN A POTENTIALLY HAZARDOUS SITUATION.
How to identify recalled test leads
There are only three Fluke test lead models affected:
- TL221 Straight to Straight
- TL222 Right to Right
- TL224 Right to Straight
The impacted leads have been depicted in the picture to the right. They have the following characteristics:
- The leads have only been available for purchase since January 2003.
- These are modular leads with no probes attached to the end.
- The connectors have a black overmolding.
- The safety rating inscribed near the logo is CAT III 1000 V, CAT IV 600 V.
- The Fluke logo can be found on the connector ends.
|
|
 |
|
Actions
- Stop using the affected TL221, TL222 and TL224 test leads.
- Verify that your test leads are affected by this recall using the picture provided.
- Return your test leads sets for free replacement:
United States residents: call 1-888-401-9940
- Please do not send your DMM, other accessories, or non-affected test leads.
|
|
|
For over 50 years, Fluke has been dedicated to the safety of our customers and we will continue to work hard to meet your high expectations for quality and safety. |
|
We at Fluke take your safety and satisfaction very seriously. Please accept our sincere apologies for this recall. If you'd like more information, please call us at 1-888-401-9940.
Best regards,

James Lico, President |