SigmaPace™ 1000 External Pacemaker Analyzer
SigmaPace™ 1000 External Pacemaker Analyzer
Overview Features Specifications Models and Accessories Knowledge and Information
Overview
Overview

The SigmaPaceTM 1000 is a powerful handheld analyzer with a comprehensive range of test suites, measurement algorithms and test loads – as specified by the external pacemaker manufacturers – to fulfill your testing requirements both quickly and efficiently.

Standard Features

  • Unique full-featured biomedical test product
  • Tests for both transcutaneous and transvenous external pacemakers
  • Full range of user-selectable measurement algorithms and test loads for external pacemakers
  • Dual-channel signal acquisition for capturing synchronous AV-sequential pulse data
  • Interactive pacemaker and ECG simulation with 5-lead output
  • Large 21-character x 8-line alphanumeric LCD readout
  • Exclusive readout "HOLD" function
  • Exclusive SigmaPaceTM 1000 test features: DC static/dynamic leakage, and current drain

Specifications
Modes of Operation
 Manual
 Remote (via standard RS232 serial port)
User Interface
Display:  - LCD readout
 - 21 characters X 8 lines
 - Brightness/viewing angle adjustment
Keys:  Eight push buttons (F-2, F-3 [UP arrow], F-4 [UP arrow], two DOWN arrows, ESCAPE, and ENTER)
Transcutaneous Pacer Tests:
Output Pulse Measurement  
Current:  - Ranges: 4.00 mA to 9.99 mA; 10.0 mA to 99.9 mA; 100 mA to 250 mA
 - Accuracy: ± 2 % of reading or ± 50 µA (whichever is greater)
Rate:  - Ranges: 5.0 PPM to 99.9 PPM; 100 PPM to 300 PPM
 - Accuracy: ± 0.5 % of reading or ± 0.3 PPM (whichever is greater)
Width:  - Ranges: 1.00 mS to 9.99 mS; 10.0 mS to 99.9 mS
 - Accuracy: ± 0.5 % of reading or ± 14 µS (whichever is greater)
Energy:  - Ranges: 1 µJ to 999 µJ; 1 mJ to 999 mJ; 1.00 J to 1.99 J
 - Accuracy: 5 % of reading/computation
Demand and Asynchronous Mode Tests  
Waveform (Physiological Simulation):  - Normal sinus rhythm (NSR)
 - Complete P-QRS-T complex
Amplitude:  1.0mVpeak (lead I)
Modes of Operation:  - Underdrive: NSR @ 85 % of pulse interval/rate
 - Overdrive: NSR @ 115 % of pulse interval/rate
 - Auxiliary Control: NSR adjustable in 1-BPM increments
 - Auxiliary Rate Range: Underdrive 10 BPM (min); overdrive 300 BPM (max)
Active Outputs:  5-lead ECG; ventricular test load; high-level ECG jack
Pacemaker Compatibility:  - Pulse Rates: 30 to 200 PPM
 - Intended Types: Demand: VVI (pace and sense); async: VOO (pace)
Amplitude Sensitivity Test  
Selections: R, S and T Waves  - Rate: 30 to 200 PPM
 - Test Loads: (30) 50 Ω to 1550 Ω in 50-Ω steps
Waveforms:  Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ)
Amplitude:  - Range: 0.05 mVpeak to 5.0 mVpeak
 - Accuracy: ± 5 % of setting
 Resolution: 0.05-mV steps (0.05 to 0.95 mVpeak); 0.50-mV steps (1 to 5 mVpeak)
Width:  - Range: 0.15 mS to 300 mS
 - Accuracy: ± 5 % of setting
 - Selections: 50
 - Resolution: 0.05-mS steps (0.15 to 0.95 mS); 1-mS steps (1 to 19 mS);
5-mS steps (20 to 95 mS)  25-mS steps (100 to 300 mS)
Active Outputs:  5-lead ECG; ventricular test load; high-level ECG jack
Pacemaker Compatibility:  - Pulse Rates: 30 to 200 PPM
 - Intended Type: VVI (pace and sense)
Noise Immunity/Line Frequency Test  
Waveform:  Sine wave
 - Frequency: 50 and 60 Hz
 - Accuracy: 0.5 Hz
Amplitude Testload Output:  - Range: 0 (OFF) to 10 mVpeak-to-peak
 - Accuracy: ± 5 % of setting
 - Resolution: 0.5-mVpeak-to-peak steps
 - Settings: (30) 50 Ω to 1550 Ω ± 1 %
5-lead ECG Output:  - Range: 0 (OFF) to 10 mVpeak-to-peak
 - Accuracy: ± 5 % of setting
 - Resolution: 0.50-mV steps
 - Reference: Lead I (RA to LA)
Active Outputs:  5-lead ECG; ventricular test load
Paced Refractory Period Test (PRP)  
Range:  20 mS to 500 mS
 - Accuracy: 5 % of reading or 1 mS (whichever is greater)
Physiological Simulation:  - Selection: Single pulse, R Wave, SSQ
 - Pulse Width: 40 mS
Outputs:  5-lead ECG; ventricular test load
Pacemaker Compatibility:  - Pulse Rates: 30 to 200 BPM
 - Intended Type: VVI (pace and sense)
Sensed Refractory Period Test (SRP)  
Range:  15 mS to 500 mS
Accuracy:  ± 5 % of reading or ± 1 mS (whichever is greater)
Physiological Simulation:  - Selection: Double pulse, R Wave, SSQ
 - Pulse Width: 40 mS
 - Amplitude: 1 mVpeak lead I
Active Outputs:  5-lead ECG; ventricular test load
Pacemaker Compatibility:  - Pulse Rates: 30 to 200 BPM
 - Intended Type: VVI (pace and sense)
Test Loads  
Transcutaneous Pacer:  - Selections: (31) 50 Ω to 1550 Ω in 50-Ω steps
 - Accuracy: ± 1 % of selection
 - Power Rating: 5 W (average); 40 W (peak) @ 1000 Ω
Input Defibrillation Protection:  - Type: Internal spark gap
 Episode Limit: 5 pulses @ 360 J (10 seconds min between discharges)
 - Life Limit: 250 pulses @ 360 J
Transvenous Pacer Tests:
Output Pulse Measurement  
Current:  - Ranges: 0.05 mA to .999 mA (available single channel only); 1.00 mA to 9.99 mA; 10 mA to 30 mA
 - Accuracy: ± 2 % of reading or ± 50 µA (whichever is greater)
 - Polarity Indicator: + or -
Rate:  - Ranges: 10.0 PPM to 99.9 PPM; 100 PPM to 999 PPM
 - Accuracy: ± 0.5 % or 0.3 PPM (whichever is greater)
Width:  - Ranges: 0.020 mS to .999 mS; 1.00 mS to 9.99 mS; 10.0 mS to 99.9 mS
 - Accuracy: 0.5 % or ± 14 µS (whichever is greater)
 - Resolution: ± 1 LSD or ± 4 µS (whichever is greater)
Voltage:  - Ranges: (available single channel only) 0.050 Vpeak to .999 Vpeak; 1.00 Vpeak to 9.99 Vpeak; 10 Vpeak to 30 Vpeak
 - Accuracy: ± 2 % of reading or ± 0.05 Vpeak (whichever is greater)
 - Polarity Indicator: + or -
Energy:  - Ranges: (available single channel only) 1 nJ to 999 nJ; 1 µJ to 999 µJ
 - Accuracy: ± 5 % of reading/computation
Display Formats:  Atrial channel only; ventricular channel only; both A + V channels
AV Interval (Delay Time)  
Measurement Ranges:  10.0 mS to 99.9 mS; 100 mS to 999 mS
Start Point:  Atrial pulse leading edge
Stop Point:  Ventricular pulse leading edge
Accuracy:  1 % of reading/computation
Demand/Async Mode Tests  
Channels:  Single and dual
Waveform:  Sine square (SSQ)
Atrial Output:  Simulated P Wave
 - Width: 30 mS
 - Amplitude: 2.0 mVpeak
Vent Output:  Simulated R Wave
 - Width: 40 mS
 - Amplitude: 2.5 mVpeak
AV Interval:  90 mS (fixed)
Interactive Simulated Rates:  - Default Settings: Underdrive = NSR @ 85 % of pulse interval/rate; overdrive = NSR @ 115 % of pulse interval/rate
 - Manual: NSR simulations adjustable in 1-BPM increments
 - Limits: Underdrive (min) = 10 BPM; overdrive (max) = 300 BPM
Output:  Ventricular channel test load; atrial channel test load
Pulse Rate:  30 to 200 PPM
Intended Pacemaker Types:  - Demand: VVI (V-channel pace and sense); AAI (A-channel pace and sense);
DDD (dual-channel pace and sense)  - Async/Continuous: VOO (V-channel pace and sense); AOO (A-channel pace and sense); DOO (dual-channel pace and sense)
Amplitude Sensitivity Test  
Operation:  Single-channel operation only (atrial or ventricular)
Atrial Channel (Physiological Simulation):  - Selection: P Wave
 - Rate: 30 to 120 BPM
 - Timing: Waveform delayed by 80 % of the pulse-to-pulse interval or 400 mS (whichever is shorter)
 - Active Output: Atrial test load
Available Test Loads:  200 Ω, 500 Ω (default setting) and 1000 Ω ± 1 %
Waveform Selections:  Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ) (default setting); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Sensitivity Waveform Amplitude:  
- Test Load Selection:  500 Ω (default setting)
 -- Range: 0.05 mVpeak to 50.0 mVpeak
 -- Accuracy: ± 5 % of setting
 --Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 50 mVpeak)
- Test Load Selection:  200 Ω
 -- Range: 0.05 mVpeak to 20.0 mVpeak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 20 mVpeak)
- Test Load Selection:  1000 Ω
 -- Range: 0.05 mVpeak to 100 mVpeak-to-peak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak); 05.0 mVpeak (50 to 100 mVpeak)
 - Default Setting: 2.0 mVpeak
Widths:  - Range: 0.15 mS to 95.0 mS
 - Accuracy: ± 5 % of setting
 - Resolution: 0.05 mS (0.15 mS to 0.95 mS); 1 mS (1 mS to 19 mS); 5 mS (20 mS to 95 mS)
Intended Pacemaker Types:  AAI (atrial pace and sense)
 - Pulse Rates: 30 to 200 PPM
Ventricular Channel (Physiological Simulation):  - Selection: R Wave, S Wave and T Wave
 - Rate: 30 to 120 BPM
 - Timing: Waveform delayed from the ventricular demand pacemaker pulse by 80 % of the pulse-to-pulse interval or 400 mS (whichever is shorter)
 - Active Output: Selected ventricular test load
Waveform Selections:  Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ) (default setting); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Available Test Load(s):  200 Ω, 500 Ω (default setting) and 1000 Ω ± 1 %
Amplitude:  
- Pacer Load Selection:  500 Ω
 -- Range: 0.05 to 50.0 mVpeak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 50 mVpeak)
- Pacer Load Selection:  200 Ω
 -- Range: 0.05 mVpeak to 20.0 mVpeak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 20 mVpeak)
- Pacer Load Selection:  1000 Ω
 -- Range: 0.05 mVpeak to 100 mVpeak-to-peak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0. 95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak); 05.0 mVpeak (50 to 100 mVpeak)
 - Default Setting: 2.5 mVpeak
Widths:  - Range: 0.15 mS to 300 mS
 - Accuracy: ± 5 % of setting
 - Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 mS to 95 mS); 25 mS (100 mS to 300 mS)
Intended Pacemaker Type(s):  VVI (atrial pace and sense only)
 - Pulse Rates: 30 to 200 PPM
Noise Immunity Test  
Channels:  Single, atrial or ventricular only
Waveform:  Sine wave
Frequency:  50 and 60 Hz
Accuracy:  ± 0.5 Hz
Active Output(s):  Atrial- or ventricular-channel test load
Output Selections:  Atrial channel only; ventricular channel only
ECG Signal:  ECG signal can be added to the selected channel.
Atrial:  1 mVpeak
Ventricular:  1.5 mVpeak
Noise Line Amplitude:  
- Pacer Load Selection:  500 Ω
 -- Range: 0 (OFF) to 100 mVpeak-to-peak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 5-mVpeak-to-peak steps
- Pacer Load Selection:  200 Ω
 -- Range: 0 (OFF) to 40 mVpeak-to-peak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 5-mVpeak-to-peak steps
- Pacer Load Selection:  1000 Ω
 -- Range: 0 (OFF) to 200 mVpeak-to-peak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 5-mVpeak-to-peak steps
Refractory Period Test (Atrial Channel)  
Test Selections:  Paced refractory period; sensed refractory period
Period:  20 to 500 mS
Accuracy:  ± 5 % of reading (or ± 1 mS, whichever is greater)
Resolution:  ± 1 LSD
Physiological Simulation:  - Selection: Square wave (default setting)
 - Atrial Channel: Simulated P Wave
 - Width: 1 mS
 - Amplitude: 20 mVpeak
 - Active Outputs: Atrial channel (4mm banana jacks) only
Additional Waveform Selections:  Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Amplitude:  - Range: .05 mVpeak to 50.0 mVpeak
 - Accuracy: ± 5 % of setting
 - Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak )
Width:  - Range: 0.15 mS to 95.0 mS
 - Accuracy: ± 5 % of setting
 - Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 to 95 mS)
Active Outputs:  Atrial channel (4mm banana jacks) only
Intended Pacemaker Types:  AAI (atrial pace and sense only)
Pacemaker Rates:  30 to 200 PPM
Available Test Load:  500 Ω ± 1 %
Refractory Period Test (Ventricular Channel)  
Test Selections:  Paced refractory period; sensed refractory period
Period:  20 to 500 mS
Accuracy:  ± 5 % of reading (or ± 1 mS, whichever is greater)
Resolution:  ± 1 LSD
Display Format:  3 digits
Physiological Simulation:  - Selection: Square wave (default setting)
 - Ventricular Channel: Simulated R Wave
 - Width: 1 mS
 - Amplitude: 20 mVpeak
 - Active Outputs: Ventricular channel (4mm banana jacks) only
Additional Waveform Selections:  Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Amplitude:  
- Pacer Load Selection:  500 Ω
 -- Range: .05 mVpeak to 50.0 mVpeak
 -- Accuracy: ± 5 % of setting
 -- Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak)
- Default setting:  20 mVpeak
Width:  - Range: 0.15 mS to 300.0 mS
 - Accuracy: ± 5 % of setting
 - Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 to 95 mS); 25 mS (100 to 300 mS)
 - Default setting: 30 mS
Intended Pacemaker Types:  VVI
Pacemaker Rates:  20 to 200 PPM
DC Leakage Current  
Measurement Range:  00.1 µA to 99.9 µA
Input Polarity:  Positive and negative
Resolution:  1 LSD (0.1 µA)
Display Format:  3 digits
Test Selections:  - Static: Continuous (power OFF)
 - Dynamic: Gated (power ON)
Test Load/Input Configurations:  - Atrial+ and atrial-
 - Ventricular+ and ventricular–
 - Atrial+ and ventricular+
Baseline/Test Selection:  500 Ω
Dynamic Test Gating Algorithm:  Measurement made 400 mS prior to the pacemaker pulse leading edge; 16 measurements averaged at a 4 mS rate for a total of 64 mS
Specified Applied Pacemaker Rate:  80 PPM
Current Drain Test  
DC Current Ranges:  0.100 mA to 0.999 mA; 1.00 mA to 9.99 mA; 10.0 mA to 99.9 mA
Polarity:  Positive or negative
Indicator:  + or – symbol
Resolution:  ± 1 LSD
Display Format:  3 digits plus decimal point
Accuracy:  ± 5 % of reading ± 10 µA
Input DC Voltage:  - Nominal: ± 9 V
 - Range: 5.0 V to 10.5 V
 - Input Protection: Short-circuit protection
 - Protection Type: Internal in-line fast-acting 1⁄2 A fuse
Selectable Testloads:  200 Ω, 500 Ω, and 1000 Ω
Battery Test Fixture:  9 V battery supply included, to facilitate connection of analyzer to recessed battery terminals within Medtronic 5388 and 5348 Temporary Pacemakers
Test Loads  
Atrial Channel:  - Selections: 200 Ω, 500 Ω, and 1000 Ω
 - Accuracy: ± 1 % of selection
 - Power Rating: 2 W
Ventricular Channel:  - Selections: 200 Ω, 500 Ω, and 1000 Ω
 - Accuracy: ± 1 % of selection
 - Power Rating: 2 W
Tracking:  Identical atrial and ventricular channel settings
Input Defibrillation Protection:  - Type: Internal spark gap
 - Episode Limit: 5 pulses @ 360 J (10 seconds min between discharges)
 - Life Limit: 250 pulses @ 360 J
Long-Term Test  
Test Configuration:  - Transvenous Pacer: Atrial or ventricular channel only
 - Transcutaneous Pacer: Ventricular channel
 - Pulse Count Range: 999,999 (max)
 - Rate: 2 % to 20 % (default setting, 10 %)
 - Amplitude: 2 % to 20 % (default setting, 10 %)
 - Test Time (max): 999:59:59 (hhh:mm:ss)
 - Maximum Error Count: 200
 - Test Termination: Manual; or upon max error count
 - Testloads: 200 Ω, 500 Ω, and 1000 Ω
Interactive Pacer ECG Simulation  
Simulation of demand, continuous, noncapture, and nonfunction patient  -ECG activity
Additional user  
-selectable parameters:  - NSR Heart Rate: Asystole and 20 to 250 BPM (1-BPM steps)
 - NSR PR Interval: 0.05 to 0.30 s (6 settings)
Pacemaker Capture/Threshold:  - Transcutaneous: 10 to 250 mA (10-mA steps)
 - Transvenous: 1 to 25 mA (1-mA steps)
General Specifications  
Serial Port:  - Type: RS232
 - Connector Type: DB-9 (male)
 - Baud Rates: 2400, 9600, and 19200
 - Data Control: Xon/Xoff
Power Requirements:  - External battery charger source/power supply
 - 100 to 240 Vac, 50/60 Hz operation
 - Auto power-off feature during battery operation
Battery Life:  20 hours (min)
Environmental Specifications  
Temperature Range:  - Operating: 15 °C to 35 °C (59 °F to 95 °F)
 - Storage: 0 °C to 50 °C (32 °F to 122 °F)
Humidity Range:  < 90 % noncondensing
Dimensions  
Size:  8” L X 4” W X 2” H (approx) (203 mm L X 101 mm W X 50 mm H)
Weight:  2 lb (approx) (0.90 kg)
CE Mark  
User Safety:  
EMC:  EN61326-1.1997
Conforms to:  UL STD 3101-1
Certified to:  CAN/USA STD C22.2 No.1010
ETL Listed  
Device has received FDA 510(k) clearance (on file)  

Model name Description Buy it **
SIGMAP1K-USA120V Sigmapace 1000 External Pacemaker Analyzer Tests both transcutaneous & transvenous external pacemakers, USA, 120 V (Item #2247700)

Includes:
  • Operator's manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2372272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 
SIGMAP1K-JPN100V Sigmapace 1000 External Pacemaker Analyzer Tests both transcutaneous & transvenous external pacemakers, Japan, 100 V (Item #2394548)

Includes:
  • Operator's manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2372272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 
SIGMAP1K-SHK250V Sigmapace 1000 External Pacemaker Analyzer Tests both transcutaneous & transvenous external pacemakers, Schuko, 250 V (Item #2394553)

Includes:
  • Operator's manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2372272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 
SIGMAP1K-UK250V Sigmapace 1000 External Pacemaker Analyzer Tests both transcutaneous & transvenous external pacemakers, United Kingdom, 250 V (Item #2394566)
Includes:
  • Operator's manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2372272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac
 
Batteries, Chargers, Power Supply Buy it **
Adapter Kit-4.0MM To Snap 9513-0202,Adapter Kit-4.0MM To Snap (for use with SigmaPace 1000) (Part#2245006)    
Powercord Australia C5 Entry 3010-0657,Powercord Australia C5 Entry (for use with SigmaPace 1000) (Part#2201443)    
Powercord Japan C5 Entry 3010-0654,Powercord Japan C5 Entry (for use with SigmaPace 1000) (Part#2201419)    
Powercord Schuko C5 Entry 3010-0656,Powercord Schuko C5 Entry (for use with SigmaPace 1000) (Part#2201437)    
Powercord UK C5 Entry 3010-0655,Powercord UK C5 Entry (for use with SigmaPace 1000) (Part#2201428)    
Powercord US C5 Entry 3010-0658,Powercord US C5 Entry (for use with SigmaPace 1000) (Part#2201455)    
Cables Buy it **
9VDC Load Test Cable 3010-0602FG, 9VDC Load Test Cable (for use with SigmaPace 1000) (Part#2392272)    
Cable 9-TO-9 Pin Null Modem 3010-0585FG, Cable 9-TO-9 Pin Null Modem (for use with SigmaPace 1000) (Part#2392260)    
Philips Agilent Cable 3010-0639,Philips Agilent Cable (for use with SigmaPace 1000) (Part#2201323)    
Miscellaneous Accessories Buy it **
HP Codemaster 3010-0605,HP Codemaster (for use with SigmaPace 1000) (Part#2201109)    
MDE/MRL R2/DAROX 3010-0607,MDE/MRL R2/DAROX (for use with SigmaPace 1000) (Part#2201127)    
MQ Responder 3010-0606,MQ Responder (for use with SigmaPace 1000) (Part#2201111)    
Physio Quik Combo 3010-0604,Physio Quik Combo (for use with SigmaPace 1000) (Part#2201095)    
Physio Quik Pace 3010-0603,Physio Quik Pace (for use with SigmaPace 1000) (Part#2201088)    
Zoll NTP 3010-0608,Zoll NTP (for use with SigmaPace 1000) (Part#2201130)    
Zoll PD & M-Series 3010-0609,Zoll PD & M-Series (for use with SigmaPace 1000) (Part#2201148)    

Infrared Camera | Infrared Detector | Thermal Imager | Thermal Imaging | Infrared Digital Camera
Predictive Maintenance

Search Fluke
ResourcesProducts
Find Resources

Home  |  Site Map  |  Fluke Corporation  |  Terms and Conditions  |  Privacy Statement  |  Disclaimer© 1995 - 2010 Fluke Corporation