Fluke
Home  |  Products  |  Calibration Instruments  |   All Calibration Instruments  |  
Contact us
Fluke 787 ProcessMeter™
Fluke 787 ProcessMeter™
Overview Features Specifications Models and Accessories Knowledge and Information
Overview
Double Your Power

A maintenance and calibration tool that will be at home in every instrumentation technician's tool box, the Fluke 787 ProcessMeter™ combines a DMM and a Loop Calibrator in one rugged, handheld tool, for about what you would expect to pay for a loop calibrator alone. Based on the trusted Fluke 87 DMM, the 787 adds the ability to measure, source, and simulate dc loop current with 0.05% accuracy and 1 microamp resolution.

  • DMM and loop calibrator in one tool
  • Clear LCD with backlight; 4,000 counts (30,000 counts for dc current)
  • Externally accessible battery for easy battery changes.
  • 1000V overload protection on V, ohms, frequency.
  • 150V overload protection on mA, backed up by 440 mA 1000V fuse.
  • 25% Manual Step plus Auto Step and Auto Ramp on mA output.
Fluke combined the DMM you trust with the loop calibrator you need.
The Fluke 787 ProcessMeter™ puts a complete troubleshooting solution in the palm of your hand. Get in touch with more information, faster and easier with the newest member of the Fluke 80 Series, the 787 ProcessMeter. The 787 gives you the measurement capabilities you expect from our rugged, reliable DMMs plus the ability to source, simulate, or measure dc current with 1 microamp resolution and accuracy of 0.05%. Put it all together and you have the ideal tool for troubleshooting and calibrating current loop applications. So if you're tired of always tracking down a calibrator, lugging around two tools and switching back and forth, get the single best solution for doubling your power. The 787 ProcessMeter from Fluke.

Here's the power the 787 ProcessMeter™ puts into your hands.



Precision 1000V, 440 mA digital multimeter
  • Measure ac and dc volts, ac and dc current, resistance, continuity, and frequency
  • 0.1% dc voltage accuracy
  • 0.05% dc current accuracy, 1 ‘A resolution to 30 mA
  • Simultaneous mA and % of scale readout
  • True rms ac voltage measurement to 1 kHz
  • Frequency measurement to 20 kHz
  • Min/Max/Average/TouchHold/Relative modes
  • Diode test and continuity beeper


DC current source/loop calibrator
  • 20 mA current source/simulator
  • Simultaneous mA and % of scale readout
  • Manual step (25%, coarse, fine) plus auto step and auto ramp


Safety Conformance

Meets 1000 volt EN61010-1 CAT III standards.

Specifications
Current Measurement
Range:  0-1 A
Resolution:  1 mA
Accuracy:  0.2% + 2 LSD
 
Range:  0-30 mA
Resolution:  0.001 mA
Accuracy:  0.05% + 2 LSD
Current Sourcing
Range:  0-20 mA or 4-20 mA
Resolution:  0.05% of span
Drive Capability:  500 Ω @ 24 mA
Frequency
 to 19.999 kHz, 0.005% + 1 LSD
Continuity
 Beeps for resistance < 100 ohms
Voltage measurement
Range:  0-1000 V AC or DC
Resolution:  0.1 mV to 1.0 V
Accuracy:  0.1% Rdg + 1 LSD (VDC)
Resistance measurement
 to 40 MΩ, 0.2% + 1 LSD
Diode test
 2.4 V shows diode voltage drop

Environmental Specifications
Operating Temperature
 -20 ºC to 55 ºC
Storage Temperature
  -40 ºC to 60 ºC
Humidity (Without Condensation)
 95% up to 30 ºC
 75% up to 40 ºC
 45% up to 50 ºC
 35% up to 55 ºC
Temperature Coefficient
 0.05 x (specified accuracy) per ºC (for temperatures < 18 ºC or > 28 ºC)

Safety Specifications
Electrical Safety
 Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92
Overvoltage category
 IEC 1010-1 CAT III
Maximum voltage
 1000 V rms

Mechanical & General Specifications
Size
 32 x 87 x 187 mm
Weight
 369 g
Batteries
 Single 9V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
Warranty
 3 years
Battery Life
 50 hours typical (measurement), 12 hours typical (sourcing 12 mA)
Shock & Vibration
 Random, 2g, 5-500 Hz
 1 meter drop test
Display current and % of span
 Yes
Auto step, auto ramp
 Yes

Model name Description Buy it **
Fluke 787 ProcessMeter

All models include:
  • AC70A Alligator Clips
  • C81Y Protective Yellow Holster with Flex-Stand™
  • One 9V alkaline battery
  • Quick Reference Guide
  • TL75 Safety-Designed Test lead Set
  • Users Manual

Pressure Buy it **
 PV350
Pressure Vacuum Module
Cases and Holsters Buy it **
 C25
Large Soft Case for DMMs
 MC6/50
MeterCleaner™ Wipes  
CXT170 Extreme Case
C100 Universal Carrying Case
C510 Leather Meter Case
C800 Hard Meter and Accessory Case
C90 Soft Case for DMMs
H80M Protective Holster with Magnetic Hanging Strap
Current Clamps Buy it **
 i1010
AC/DC Current Clamp
 i410
AC/DC Current Clamp
i200 AC Current Clamp
Fiber Optics Buy it **
 FOM
Fiber Optic Meter
Temperature Buy it **
 80TK
Thermocouple Module (Type K)
80PK-22 Immersion Temperature Probe
80PK-24 Air Temperature Probe
80PK-25 Piercing Temperature Probe
80PK-26 Tapered Temperature Probe
80PK-27 Industrial Surface Temperature Probe
Discontinued Products Buy it **
 C81Y
Meter Holster (Yellow)    
TL74 Test Lead Set (4 mm)    
80i-400 AC Current Clamp (400 A)    
TL70A Precision Probe Assembly    
C28Y/WWG Carrying Case, Yellow for WWG    
C28Y Durable Case    
C81G Meter Holster (Grey)    
Search Fluke
ResourcesProducts
Find Resources

Home  |  Site Map  |  Fluke Corporation  |  Terms and Conditions  |  Privacy Statement  |  Disclaimer© 1995 - 2010 Fluke Corporation