The Fluke 718 pressure calibrator family (includes 4 models) provide a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.
Features
Pressure Source and Milliamp Measurement to calibrate and maintain almost any pressure device
Integrated pump is easily cleaned when accidently exposed to fluids that reduces cost of ownership and repairs and enables servicing the pump in the field
Error % calculation for quicker pass/fail decision
making in the field
Best in class mA accuracy of 0.015 % for better
measurement confidence and workload coverage
Min/Max/Hold to capture changing measurements
Switch test feature for quicker and more accurate
pressure switch testing
New 1 and 300 PSI pressure ranges mean fewer
extra tools required
1, 30, 100 & 300 PSI ranges now available
Specifications
Technical Specifications
Standard
718 Series
Measure Pressure
(internal sensor)
Accuracy
0.05% full scale
Functions
zero, min/max, hold, damp
Note:
Dry air, non-corrosives only
718 1G
Range:
-1 PSI to +1 PSI, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa)
Resolution:
0.0001 psi, 0.001 mbar, 0.0001 kPa
Over Pressure:
5x Full Scale
Functions:
Zero, Min, Max, Hold, Damp
718 30G
Range:
-12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa)
Resolution:
0.001 psi, 0.1 mbar, 0.01 kPa
Over Pressure:
5x Full Scale
Functions:
Zero, Min, Max, Hold, Damp
718 100G
Range:
-12 PSI to 100 PSI, (-850 mbar to 7 bar, -85 to 689.48 kPa)
Resolution:
0.01 psi, 0.1 mbar, 0.01 kPa
Over Pressure:
2x Full Scale
Functions:
Zero, Min, Max, Hold, Damp
718 300G
Range:
-12 PSI to 300 PSI, (-850 mbar to 20 bar, -85 to 2068.42 kPa )
Resolution:
0.01 psi, 1 mbar, 0.1 kPa
Over Pressure:
375 PSI 25 bar
Functions:
Zero, Min, Max, Hold, Damp
Fluke 718 series via using 700 Series Pressure Modules
Range:
29 pressure modules, 0 – 1 in H20 to 10,000 psi, 2.5 mbar to
700 bar. Over pressure per pressure module specs
Resolution:
Per pressure module specs
Accuracy:
To 0.025 % of full span, per pressure module specs
Functions:
Zero, Min, Max, Hold, Damp
Note: media compatibility per pressure module specs
Fluke 718 series via built in pump
Range:
-12 PSI or -850 mbar to full scale Supported Pressure Units; psi, in H20 (4 °C), in
H20 (20 °C), cm H20 (4 °C), cm H20 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop Power:
Range: 24 V dc
Accuracy: +/- 10 %
Note: Drive; 20 mA into 1000 Ω for battery > 6.8V; 700 Ω
for battery 5.8 to 6.8 V
mA Measure:
Range: 0 mA to 24 mA
Resolution: 0.001 mA
Accuracy: 0.015 % + 1 count
General Specifications
Maximum voltage
30 volts: Non-operating
Temperature
-40 °C to 60 °C
Operating Temperature
-10 °C to 55° C
Relative Humidity
95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C);
45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating Altitude
3,000 M max
Shock
1 M drop test
Vibration
Random, 2 g, 5 Hz – 500 Hz
Safety
CSA C22.2 No. 1010.1: 1992
EMC
EN50082-1: 1992 and EN55022: 1994 Class B
Size
216 mm x 94 mm x 66 mm
(8.50 in x 3.72 in x 2.60 in)
Weight
992 g (35 oz)
Power, battery life
9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two
batteries in 718
Battery Life
4 to 20 hours typical, depending on functions used
Warranty
Three years (one year on pressure pump in Fluke 718)
Display
LCD, 5 digit pressure and current simultaneous
Note
For use with non-corrosive gasses
Models and Accessories
Model name
Description
Buy it **
Fluke 718 1G
Pressure Calibrator -1 to 1 PSI, (-68.9 mbar to 68.9 mbar, -6.89 to 6.89 kPa)
All models include:
Protective Yellow Holster with test lead storage
Instructions Sheet (14 languages)
Test Leads and Alligator Clip
Two 9V alkaline batteries
NIST - Traceable calibration report and data
Fluke 718 30G
Pressure Calibrator -12 PSI to 30 PSI, (-850 mbar to
2 bar, -85 to 206.84 kPa)
Fluke 718 100G
Pressure Calibrator -12 PSI to 100 PSI, (-850 mbar to
7 bar, -85 to 689.48 kPa)
Fluke 718 300G
Pressure Calibrator -12 PSI to 300 PSI, (-850 mbar to
20 bar, -85 to 2068.42 kPa )