Fluke
Home  |  Products  |  Process Calibration Tools  |   Loop Calibrators  |  
Contact us
Fluke 707Ex Intrinsically safe loop calibrator

 

Fluke 707Ex Intrinsically safe loop calibrator
Overview Features Specifications Models and Accessories Knowledge and Information
Overview
The fast one-handed tool for mA loop checks in Ex zones

The 707Ex is an intrinsically safe loop calibrator for use in explosion endangered areas. The Fluke 707Ex is certified in accordance with the ATEX directive (Ex II 2 G Eex ia IIC T4) in Zones 1 and 2.

  • Large display and simple, quick-click push rotary button for easy one-handed operation
  • Simultaneous mA and % readout for quick, easy, interpretation of readings.
  • mA accuracy of 0.015%, superior to other loop calibrators
  • 1 µA resolution for mA source, simulate and measure
  • Push button with 25% steps for fast, easy linearity checks
  • 0-100% "span check" for fast confirmation of zero and span
  • Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests
  • Internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM
  • Measures Vdc to 28 V
  • Single 9 V battery that is easily changed
  • 0-20 mA or 4-20 mA default start up modes
  • Innovative output adjustment dial with 1 µA and 100 µA resolution
  • HART® compatible resistance is connected in series with the loop supply for compatibility with HART communicators
Specifications
Current Measurement
Range:  0 to 24 mA
Resolution:  0.001 mA
Accuracy:  0.015% Rdg + 2 LSD
Current Sourcing
Range:  0 to 20 mA or 4 to 20 mA
Accuracy:  0.015% Rdg + 2 LSD
Drive Capability:  700 Ω at 20 mA
Current Simulation
Range:  0 to 20 mA or 4 to 20 mA
Accuracy:  0.015% Rdg + 2 LSD
Loop Power
 While measuring mA, 24 V
Voltage measurement
Range:  0 to 28 V DC
Resolution:  1 mV
Accuracy:  0.015% Rdg + 2 LSD

Environmental Specifications
Operating Temperature
 -10ºC to 55ºC
Storage Temperature
 -40ºC to 60ºC
Humidity (Without Condensation)
 0 to 95% (0ºC to 30ºC);
 0 to 75% (30ºC to 40ºC)
 0 to 45% (40ºC to 50ºC)
Temperature Coefficient
 -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC

Safety Specifications
Safety
 CSA C22.2 No. 1010.1:1992
EMC
 EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage
 30 V
ATEX certification
 Ex II G Eex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X
CE certification
 EN61010-1 and EN61326

Mechanical & General Specifications
Size
 152 x 70 x 41 mm
Weight
 224 g (350 g including holster)
Batteries
 One 9V battery Alkaline
Warranty
 3 year
Shock & Vibration
 per MIL-T-28800 for a Class 2 Instrument
Display current and % of span
 Yes
Auto step, auto ramp
 Yes
Span Check
 Yes

Model name Description Buy it **
Fluke 707Ex Intrinsically safe loop calibrator
  • Alligator test clips
  • Instruction manual (on CD-ROM)
  • Protective holster with Flex-Stand™
  • Safety designed test leads
  • NIST- traceable calibration report and data
Pressure Buy it **
PV350 Pressure Vacuum Module
Cases and Holsters Buy it **
C10 Meter Holster
C100 Universal Carrying Case
C12A Soft Meter Case
C195 Soft Case    
C25 Large Soft Case for DMMs
C800 Hard Meter and Accessory Case
Temperature Buy it **
80T-150UA Universal Temperature Probe (for DMMs)
80TK Thermocouple Module (Type K)
Discontinued Products Buy it **
80T-IR Infrared Temperature Probe (for DMMs)    
Search Fluke
ResourcesProducts
Find Resources

Home  |  Site Map  |  Fluke Corporation  |  Terms and Conditions  |  Privacy Statement  |  Disclaimer© 1995 - 2010 Fluke Corporation