The fast one-handed tool for mA loop checks in Ex zones
The 707Ex is an intrinsically safe loop calibrator for use in explosion endangered areas. The Fluke 707Ex is certified in accordance with the ATEX directive (Ex II 2 G Eex ia IIC T4) in Zones 1 and 2.
Features
Large display and simple, quick-click push rotary button for easy one-handed operation
Simultaneous mA and % readout for quick, easy, interpretation of readings.
mA accuracy of 0.015%, superior to other loop calibrators
1 µA resolution for mA source, simulate and measure
Push button with 25% steps for fast, easy linearity checks
0-100% "span check" for fast confirmation of zero and span
Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests
Internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM
Measures Vdc to 28 V
Single 9 V battery that is easily changed
0-20 mA or 4-20 mA default start up modes
Innovative output adjustment dial with 1 µA and 100 µA resolution
HART® compatible resistance is connected in series with the loop supply for compatibility with HART communicators
Specifications
Specifications
Current Measurement
Range:
0 to 24 mA
Resolution:
0.001 mA
Accuracy:
0.015% Rdg + 2 LSD
Current Sourcing
Range:
0 to 20 mA or 4 to 20 mA
Accuracy:
0.015% Rdg + 2 LSD
Drive Capability:
700 Ω at 20 mA
Current Simulation
Range:
0 to 20 mA or 4 to 20 mA
Accuracy:
0.015% Rdg + 2 LSD
Loop Power
While measuring mA, 24 V
Voltage measurement
Range:
0 to 28 V DC
Resolution:
1 mV
Accuracy:
0.015% Rdg + 2 LSD
Environmental Specifications
Operating Temperature
-10ºC to 55ºC
Storage Temperature
-40ºC to 60ºC
Humidity (Without Condensation)
0 to 95% (0ºC to 30ºC);
0 to 75% (30ºC to 40ºC)
0 to 45% (40ºC to 50ºC)
Temperature Coefficient
-10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC
Safety Specifications
Safety
CSA C22.2 No. 1010.1:1992
EMC
EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage
30 V
ATEX certification
Ex II G Eex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X