| Specifications |
| Resistance | |
| Temperature coefficient | |
| Temperature range |
|
-200 C to 300 ºC (transition and cable temperature: 0 ºC to 150 ºC) |
|
| Drift rate |
|
+ 0.13 ºC at 0 ºC after 1000 hours at 300 ºC |
|
| Sheath Material | |
| Leads |
|
Teflon-insulated, nickel-plated stranded copper, 22 AWG |
|
| Termination | |
| Time constant |
|
Four seconds maximum for 63.2% response to step change in water moving at 3 fps |
|
| Bending radius |
|
19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip |
|
| Calibration |
|
Includes manufacturer’s NIST-traceable calibration and table with R vs. T values in 1 ºC increments from -196 ºC to 300 ºC. Callendar-van Dusen coefficients included |
|
| Immersion |
|
At least 100 mm (4 in) recommended |
|
| Accuracy (includes calibration uncertainty and short-term stability) |
|
±0.50 ºC at -196 ºC |
|
±0.050 ºC at 0 ºC |
|
±0.052 ºC at 200 ºC |
|
±0.055 ºC at 300 ºC |
|
| Size |
|
9 in L x 3/16 in diameter |
|
| Marking |
|
Each RTD is marked with the Fluke name and part number |
|