“Simply Powerful!”
Take along the new Fluke 725 Multifunction Process Calibrator, and you're equipped to test and calibrate almost any process parameter. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules. Do you need to calibrate transmitters? The 725's split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725 has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests. From the moment you pick it up, the Fluke 725 is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.
|
The Fluke 725 Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.
- Small, streamlined shape makes it easy to carry
- Rugged, reliable design stands up to field use
- Easy to read measure/source screen lets you view input and output simultaneously
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
- Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
- Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
- Source mA with simultaneous pressure measurement to conduct valve and I/P tests
- Support flow meter testing with frequency and CPM functions
- Perform fast linearity tests with auto- step and auto-ramp features
- Power transmitters during test using loop supply with simultaneous mA measurement
- Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
- Store frequently used test setups for later use
- Backlight lets you work in poor light
- Large battery capacity of four AA cells
- Battery door for easy changes
|
|
| Measurement Accuracy |
| Voltage DC |
| 30.000 V |
0.02%+ 2 counts (upper display) |
| 30.000 V |
0.02%+ 2 counts (lower display) |
| 100.00 mV |
0.02%+ 2 counts |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
| | Current DC |
| 24.000 mA |
0.02%+ 2 counts |
| | Resistance |
| 0.0 to 400.0 Ω |
0.1Ω (4-wire), 0.15Ω (2- and 3-wire) |
| 401 to 1500 Ω |
0.5Ω (4-wire), 1Ω (2- and 3-wire) |
| 1500 to 3200 Ω |
1Ω (4-wire), 1.5Ω (2- and 3-wire) |
| | Frequency |
| 2.0 to 1000.0 CPM |
0.05 % + 1 count |
| 1.0 to 1100.0 Hz |
0.05 % + 1 count |
| 1.00 to 10.00 kHz |
0.05 % + 1 count |
| Sensitivity |
1 V peak-to-peak-minimum |
| | Pressure |
| Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
|
| Source Accuracy |
| Voltage DC |
| 100.00 mV |
0.02%+2 counts |
| 10.000 V |
0.02%+2 counts |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
| | Current DC |
| 24.000 mA (Source) |
0.02%+ 2 counts |
| 24.000 mA (Simulate) |
0.02%+ 2 counts |
| | Resistance |
| 15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω |
0.5 Ω(excitation current 0.05 to 0.8 mA) |
| 1500 to 3200 Ω |
1 Ω(excitation current 0.05 to 0.4 mA) |
| | Frequency |
| 2.0 to 1000.0 CPM |
0.05% |
| 1.0 to 1100.0 Hz |
0.05% |
| 1.00 to 10.00 kHz |
0.25% |
| Waveform: |
5 V p-p squarewave, -0.1 V offset |
|
| RTDs and Thermocouples |
| Measure Accuracy: |
| NI-120 |
0.2 °C |
| PT-100 (385) |
0.33 °C |
| PT-100 (393) |
0.3 °C |
| PT-100 (JIS) |
0.3 °C |
| PT-200 (385) |
0.2 °C |
| PT-500 (385) |
0.3 °C |
| PT-1000 (385) |
0.2 °C |
| Resolution: |
0.1 °C |
| J: |
0.7 °C |
| K: |
0.8 °C |
| T: |
0.8 °C |
| E: |
0.7 °C |
| R: |
1.8 °C |
| S: |
1.5 °C |
| B: |
1.4 °C |
| L: |
0.7 °C |
| U: |
0.75 °C |
| N: |
0.9 °C |
| Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
| XK |
0.6°C |
| BP |
1.2°C |
| | Source Accuracy: |
| NI-120 |
0.2 °C |
| PT-100 (385) |
0.33 °C |
| PT-100 (393) |
0.3 °C |
| PT-100 (JIS) |
0.3 °C |
| PT-200 (385) |
0.2 °C |
| PT-500 (385) |
0.3 °C |
| PT-1000 (385) |
0.2 °C |
| Resolution: |
0.1 °C |
| Note: |
Accuracy stated for 4-wire measurement. |
| J: |
0.7 °C |
| K: |
0.8 °C |
| T: |
0.8 °C |
| E: |
0.7 °C |
| R: |
1.4 °C |
| S: |
1.5 °C |
| B: |
1.4 °C |
| L: |
0.7 °C |
| U: |
0.75 °C |
| N: |
0.9 °C |
| Resolution |
J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C |
| XK |
0.6°C |
| BP |
1.2°C |
|
| Specifications |
| Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
| | Loop Power Function |
| Voltage: |
24 V |
| Accuracy: |
10% |
| Maximum current: |
22 mA, short circuit protected |
| | Step Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Steps |
25% of range, 100% of range |
|
| Environmental Specifications |
| Operating Temperature | | | Storage Temperature | | | Operating Altitude | |
| Safety Specifications |
| Agency Approvals |
|
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
|
| Mechanical & General Specifications |
| Size | | | Weight | | | Batteries | | | Battery Replacement |
|
Separate battery compartment, accessable without breaking calibration seal |
| | Side port connections |
|
Pressure module connector, also used for remote real-time programming |
|
|
| Model name | Description | Buy it ** | | Fluke 725 | Multifunction Process Calibrator including:
- One set of AC72 alligator clips
- One set of stackable test leads
- One set of TL 75 testleads
- User manuals
- NIST- traceable calibration report and data
|
|  |
| Differential pressure modules | Buy it ** | | Differential Pressure Module (2.5 mbar) | | | | Differential Pressure Module (70 mbar) | | | | Differential Pressure Module (340 mbar) | | | | Differential Pressure Module (1000 mbar) | | | | 700P01 | Differential Pressure Module (25 mbar) | | |
| Pressure Calibration Kit | Buy it ** | | 700PCK | Pressure Module Calibration Kit | | |
| Pumps and Accessories | Buy it ** | | 700PMP | Pressure Pump (10 bar) | |  |
| Test leads, Probes and Clips | Buy it ** | | SureGrip™ Alligator Clips |  |  | | Alligator Clip Set |  |  | | SureGrip™ Industrial Test Lead Set |  |  | | Test Lead Set for High Energy Environments |  |  | | Test Lead Set |  |  | | AC280 | SureGrip™ Hook Clips |  |  | | AC283 | SureGrip™ Pincer Clips |  |  | | AC285 | SureGrip™ Alligator Clips |  |  | | AC87 | Heavy Duty Bus Bar Clips |  |  | | AC89 | Heavy Duty Insulation Piercing Test Clip |  |  | | BP980 | Double Banana Plug Kit |  |  | | H900 | Test Lead Holder |  |  | | TL221 | SureGrip™ Test Lead Extension Set |  |  | | TL222 | SureGrip™Silicone Insulated Test Leads |  |  | | TL223 | SureGrip™ Electrical Test Lead Set |  |  | | TL224 | SureGrip™ Insulated Test Leads |  |  | | TL26A | 60 in. 5-Way Test Lead Set |  |  | | TL27 | Heavy Duty Test Lead Set |  |  | | TL28A | Automotive Test Lead Set |  |  | | TL40 | Retractable Tip Test Lead Set |  |  | | TL75 | Hard Point Test Lead Set |  |  | | TL76 | 2mm/4mm Test Lead Set |  |  | | TL80A | Basic Electronic Test Lead Set | |  | | TL81A | Deluxe Electronic Test Lead Set |  |  | | TL910 | Electronic Test Probes |  |  | | TL930 | Patch Cords (60 cm) |  |  | | TL932 | Patch Cords (90 cm) |  |  | | TL935 | Patch Cords (60, 90, 120 cm) |  |  | | TL940 | Mini-Hook Test Leads |  |  | | TL950 | Mini-Pincer Test Leads |  |  | | TL960 | Micro-Hook Test Leads |  |  | | TL970 | Hook and Pincer Set |  |  | | TP1 | Slim Reach Test Probes (flat blade) |  |  | | TP2 | Slim Reach Test Probes (2 mm) |  |  | | TP220 | SureGrip™ Industrial Test Probes |  |  | | TP38 | Slim Reach Test Probe (insulated) |  |  | | TP4 | Slim Reach Test Probes (4 mm) |  |  | | TP74 | Lantern Tip Test Probes |  |  | | TP80 | Electronic Test Probes |  |  | | TP912 | Replacement Tips for TL910 |  |  | | TP920 | Test Probe Adapter Set |  |  |
| Gage pressure modules | Buy it ** | | Gage Pressure Module (34 bar) | | | | Gage Pressure Module (70 bar) | | | | Gage Pressure Module (340 bar) | | | | Gage Pressure Module (700 bar) | | | | 700P05 | Gage Pressure Module (2 bar) | | | | 700P06 | Gage Pressure Module (7 bar) | | | | 700P09 | Gage Pressure Module (100 bar) | | | | 700P27 | Gage Pressure Module (20 bar) | | | | 700P29 | Gage Pressure Module (200 bar) | | |
| Cases-Holsters | Buy it ** | | Soft Meter Case | | | | LockPak, Meter Locking Kit | | | | ToolPak, Meter Hanging Kit |  |  | | C100 | Universal Carrying Case |  |  | | C120 | Hard Carrying Case | |  | | C25 | Large Soft Case for DMMs |  |  | | C510 | Leather Meter Case |  |  | | C550 | Tool Bag |  |  | | C75 | Soft Accessory Case |  |  | | C781 | Soft Meter Case |  |  | | CXT80 | Extreme Case |  |  |
| Differential pressure module wet | Buy it ** | | Differential Pressure Module Wet (70 mbar) | | | | Differential Pressure Module Wet (340 mbar) | | | | 700P24 | Differential Pressure Module Wet (1000 mbar) | | |
| Absolute pressure modules | Buy it ** | | Absolute Pressure Module (340 mbar) | | | | Absolute Pressure Module (2 bar) | | | | Absolute Pressure Module (7 bar) | | | | 700PA4 | Gage Pressure Module (1000 mbar) | | |
| Current-Clamps | Buy it ** | | AC/DC Current Clamp |  |  | | AC/DC Current Clamp |  |  |
| Dual pressure modules | Buy it ** | | Dual Pressure Module (± 70 mbar) | | | | Dual Pressure Module (± 340 mbar) | | | | Dual Pressure Module (±1000 mbar) | | | | Dual Pressure Module (-1000/+2000 mbar) | | | | Dual Pressure Module (-1000 mbar/+6.9 bar) | | | | Dual Pressure Module ( -1000 mbar/+13.8 bar) | | |
|
|
|
|
 |
 | | Search Fluke | | | | | | | | | | | | | | | | | | | | | | | | Find Resources | | | | | | | | |
|
|
 |
|
|
|
|
|
|
 |
| | |