Do more, carry less
The 743B does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. For documentation, the 743B automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. The 743B has a PC interface that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. With its expanded memory, the 743B can hold a full week of calibrations and procedures.
|
The 743B is easy to use with flexible test strategies. And like all Fluke tools, it's rugged, reliable and tough as nails. It's like having several powerful tools in one - so you can work faster than ever:
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
- Power transmitters during test using loop supply with simultaneous mA measurement
- Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
- Measure and source with one compact, rugged, reliable tool
- Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
- Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
- Easy to use
- Three year warranty
- Bright white display. Read both sourced and measured parameters.
- Multi lingual interface
- Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1msec.
- Standard delivered with DPC/Track Sample software. Compatible with many Asset Management software packages.
|
|
| Measurement Accuracy |
| Voltage DC |
| 110.000 mV |
0.025%+0.015% * |
| 1.10000 V |
0.025%+0.005% * |
| 11.0000 V |
0.025%+0.005% * |
| 110.000 V |
0.05%+0.005% * |
| 300.00 V |
0.05%+0.005% * |
| | Voltage AC |
| 20 to 40 Hz |
2% + 10 |
| 40 to 500 Hz |
0.5% + 5 |
| 500 to 1 kHz |
2% + 10 |
| 1 kHz to 5 kHz |
10% + 20 |
| Ranges: |
1.1000, 11.000, 110.00, 300V |
| | Current DC |
| 30.000 mA |
0.01% + 0.015% * |
| 110.00 mA |
0.01% + 0.015% * |
| | Resistance |
| 11.000 Ω |
0.05% + 50 mΩ |
| 110.00 Ω |
0.05% + 50 mΩ |
| 1.1000 kΩ |
0.05% + 500 mΩ |
| 11.000 kΩ |
0.1% + 10 Ω |
| | Frequency |
| 1.00 to 109.99 Hz |
0.05 Hz |
| 110.0 to 1099.9 Hz |
0.5 Hz |
| 1.100 to 10.999 kHz |
5 Hz |
| 11.00 to 50.00 kHz |
50 Hz |
| | Pressure |
| Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
| | Note |
| * |
(% of reading + % of full scale) |
|
| Source Accuracy |
| Voltage DC |
| 110.000 mV |
0.01%+0.005% * |
| 1.10000 V |
0.01%+0.005% * |
| 15.0000 V |
0.01%+0.005% * |
| | Current DC |
| 22.000 mA (Source) |
0.01%+ 0.015% * |
| 22.000 mA (Simulate) |
0.02% + 0.03% * |
| | Resistance |
| 11.000 Ω |
0.01% + 20 mΩ |
| 110.00 Ω |
0.01% + 40 mΩ |
| 1.1000 kΩ |
0.02% + 500 mΩ |
| 11.000 kΩ |
0.03% + 5Ω |
| | Frequency |
| 0.00 to 10.99 Hz |
0.01 Hz |
| 11.00 to 109.99 Hz |
0.1 Hz |
| 110.0 to 1099.9 Hz |
0.1 Hz |
| 1100 to 21999 Hz |
2 Hz |
| 22.000 to 50.000 kHz |
5 Hz |
| | Note |
| * |
(% of reading + % of full scale) |
|
| RTDs and Thermocouples |
| Measure Accuracy: |
| 10 Ω Cu (427):
|
2 °C |
| 100 Ω Pt (3916): |
0.3 °C |
| 100 Ω Pt (3926): |
0.3 °C |
| 100 Ω Pt (385): |
0.3 °C |
| 200 Ω Pt (385): |
0.3 °C |
| 500 Ω Pt (385): |
0.3 °C |
| 1000 Ω Pt (385): |
0.3 °C |
| 120 Ω Ni (672): |
0.3 °C |
| note: |
For 2-and 3-wire measurement, add 0.4 °C |
| E: |
0.3 °C |
| N: |
0.5 °C |
| J: |
0.3 °C |
| L: |
0.3 °C |
| K: |
0.3 °C |
| T: |
0.3 °C |
| U: |
0.3 °C |
| B: |
0.9 °C |
| R: |
1.0 °C |
| S: |
0.9 °C |
| C: |
0.6 °C |
| note: |
Accuracy with external cold junction, for internal junction add 0.2 °C |
| | Source Accuracy: |
| 10 Ω Cu (427): |
1 °C |
| 100 Ω Pt (3916): |
0.1°C |
| 100 Ω Pt (3926): |
0.1°C |
| 100 Ω Pt (385): |
0.1°C |
| 200 Ω Pt (385): |
0.1°C |
| 500 Ω Pt (385): |
0.1°C |
| 1000 Ω Pt (385): |
0.1°C |
| 120 Ω Ni (672): |
0.1°C |
| note: |
For 2-and 3-wire simulation, add 0.4 °C |
| E: |
0.2 °C |
| N: |
0.3 °C |
| J: |
0.2 °C |
| L: |
0.2 °C |
| K: |
0.3 °C |
| T: |
0.3 °C |
| U: |
0.3 °C |
| B: |
0.8 °C |
| R: |
0.9 °C |
| S: |
0.9 °C |
| C: |
0.6 °C |
| note: |
Accuracy with external cold junction, for internal junction add 0.2 °C |
|
| Specifications |
| Data Log Functions |
| Measure functions: |
Voltage, current, resistance, frequency, temperature, pressure |
| Reading rate: |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
| Maximum record length: |
8000 readings (7980 for 30 or 60 readings/minute) |
| | Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Rate: |
4 steps/second |
| Trip detect: |
Continuity or voltage (continuity detection not available when sourcing current) |
| | Loop Power Function |
| Voltage: |
Selectable, 24 V or 28 V |
| Accuracy: |
5% |
| Maximum current: |
22 mA, short circuit protected |
| Maximum input voltage: |
30 V DC |
| | Step Functions |
| Source Functions |
Voltage, current, resistance, frequency, temperature |
| Manual Step |
Selectable step, change with arrow buttons |
| Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
|
| Environmental Specifications |
| Operating Temperature |
|
-10°C to +50°C |
|
-20°C to +50°C * |
| | Storage Temperature | | | Dust/water resistance | | | Operating Altitude | | | Note |
| * |
Except frequency and AC |
|
| Safety Specifications |
| Agency Approvals |
|
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
|
| Mechanical & General Specifications |
| Size | | | Weight | | | Batteries | | | Battery Life | | | Battery Replacement |
|
Via snap-shut door without opening calibrator; no tools required |
| | Side port connections |
|
Pressure module connector |
|
RS-232 connector for PC interface cable |
|
Connection for optional battery eliminator
|
| | Data storage capacity |
|
1 week of calibration results |
| | 90 day specifications |
|
The standard specification intervals for the 743B is 1 and 2 years. |
|
Typical 90 day measurement and source accuracy can be estimated by |
|
dividing the one year "% of reading" or "%of output" specifications by 2. |
|
Floor specifications, expressed as "% of full scale" or |
|
"counts" or "ohms" remain constant. |
|
|
| Model name | Description | Buy it ** | | Fluke 743B | Documenting Process Calibrator including:
- BC7217 battery charger
- BP7217 battery pack
- DPC/TRACK Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- One set of TP220 test probes
- Serial port cable
- Two sets of AC220 test clips
- Two sets of TL224 industrial test leads
|
|  |
| Differential pressure modules | Buy it ** | | 700P00 | Differential Pressure Module 1 in. H2O (2.5 mbar) | | | | 700P01 | Differential Pressure Module 10 in H2O (25 mbar) | | | | 700P02 | Differential Pressure Module 1 psi (70 mbar) | | | | 700P03 | Differential Pressure Module 5 psi (340 mbar) | | | | 700P04 | Differential Pressure Module 15 psi (1000 mbar) | | |
| Pressure Calibration Kit | Buy it ** | | 700PCK | Pressure Module Calibration Kit | | |
| Pumps and Accessories | Buy it ** | | 700HTH-1 | Hydraulic Test Hose | | | | 700HTP-1 | Hydraulic Test Pump | |  | | 700PMP | Pressure Pump 150 psi (10 bar) | |  | | 700PRV-1 | Pressure Relief Valve Kit | |  | | 700PTP-1 | Pneumatic Test Pump | |  |
| Test leads Probes and Clips | Buy it ** | | SureGrip™ Alligator Clips |  |  | | AC280 | SureGrip™ Hook Clips |  |  | | AC283 | SureGrip™ Pincer Clips |  |  | | AC285 | SureGrip™ Alligator Clips |  |  | | TL940 | Mini-Hook Test Leads |  |  | | TL950 | Mini-Pincer Test Leads |  |  | | TP920 | Test Probe Adapter Set |  |  |
| Gauge pressure modules | Buy it ** | | 700P05 | Gauge Pressure Module 30 psi (2 bar) | | | | 700P06 | Gauge Pressure Module 100 psi (7 bar) | | | | 700P07 | Gauge Pressure Module 500 psi (34 bar) | | | | 700P08 | Gauge Pressure Module 1000 psi (70 bar) | | | | 700P09 | Gauge Pressure Module 1500 psi (100 bar) | | | | 700P27 | Gauge Pressure Module 300 psi (20 bar) | | | | 700P29 | Gauge Pressure Module 3000 psi (200 bar) | | | | 700P30 | Gauge Pressure Module 5000 psi (340 bar) | | | | 700P31 | Gauge Pressure Module 10000 psi (700 bar) | | |
| Cases-Holsters | Buy it ** | | Soft Meter Case |  |  | | C550 | Tool Bag |  |  | | C700 | Hard Case (700 Series) | |  | | C75 | Soft Accessory Case |  |  | | C781 | Soft Meter Case |  |  |
| Differential pressure module wet | Buy it ** | | 700P22 | Differential Pressure Module Wet 1 psi (70 mbar) | | | | 700P23 | Differential Pressure Module Wet 5 psi (340 mbar) | | | | 700P24 | Differential Pressure Module Wet 15 psi (1000 mbar) | | |
| Absolute pressure modules | Buy it ** | | Pneumatic Test Pump (0.8 bar/40 bar) | | | | 700HTP-1 | Hydraulic Test Pump (700 bar) | | | | 700PA3 | Absolute Pressure Module 5 psi (340 mbar) | | | | 700PA4 | Gauge Pressure Module 15 psi (1000 mbar) | | | | 700PA5 | Absolute Pressure Module 30 psi (2 bar) | | | | 700PA6 | Absolute Pressure Module 100 psi (7 bar) | | | | 700PRV-1 | Pressure Relief Valve Kit | | |
| Current-Clamps | Buy it ** | | i200 | AC Current Clamp |  |  | | i200s | AC Current Clamp |  |  |
| Dual pressure modules | Buy it ** | | 700PD2 | Dual Pressure Module ± 1 psi (± 70 mbar) | | | | 700PD3 | Dual Pressure Module ± 5 psi (± 340 mbar) | | | | 700PD4 | Dual Pressure Module ± 15 psi (±1000 mbar) | | | | 700PD5 | Dual Pressure Module -15/30 PSI (-1000/+2000 mbar) | | | | 700PD6 | Dual Pressure Module -15/1000 psi (-1000 mbar/+6.9 bar) | | | | 700PD7 | Dual Pressure Module -15/200 psi ( -1000 mbar/+13.8 bar) | | |
| Vacuum pressure modules | Buy it ** | | 700PV3 | Vacuum Pressure Module -5 psi (-340 mbar) | | | | 700PV4 | Vacuum Pressure Module -15 psi (-1000 mbar) | | |
| Fiber-Optics | Buy it ** | | FOA ST/ST | Fiber Optic Coupler | | | | FOC-ST/SMA | Fiber Optic Patch Cords | | |
| Temperature | Buy it ** | | Thermocouple Plug Kit (5 types) | |  | | Thermocouple Plug Kit (10 types) | |  | | 80CJ-M | Male Mini Connectors (Type J) |  |  | | 80CK-M | Male Mini Connectors (Type K) |  |  | | 80PJ-EXT | Extension Wire Kit (Type J) |  |  | | 80PK-1 | Bead Probe (Type K) |  |  | | 80PK-22 | Immersion Temperature Probe |  |  | | 80PK-24 | Air Temperature Probe |  |  | | 80PK-25 | Piercing Temperature Probe |  |  | | 80PK-26 | Tapered Temperature Probe |  |  | | 80PK-27 | Industrial Surface Temperature Probe |  |  | | 80PK-8 | Pipe Clamp Temperature Probe (Type K) |  |  | | 80PK-9 | General Purpose Probe (Type K) |  |  | | 80PK-EXT | Extension Wire Kit (Type K) |  |  | | 80PT-25 | Piercing Temperature Probe |  |  |
|
|
|
|
 |
 | | Search Fluke | | | | | | | | | | | | | | | | | | | | | | | | Find Resources | | | | | | | | |
|
|
 |
|
|
|
|
|
|
|
 |
| | |